FAQ


Q&A

  • Why ultrasonic?
  • How to purchase?
  • More about sample testing
 
  • More about explosion proof model
  • More about demo trial

Concentration Meter (FUD-1 series)

  • How to calibrate?
  • How to check the condition?
  • Any replacement item?
 
  • Any caution at installation?
  • Measurement errors

Viscosity Meter (FUV-1 series)

  • How to calibrate?
  • How to check the condition?
  • Any replacement item?
 
  • Measurement range?

Desktop Viscosity Meter (FCV-100)

  • Measurement range?
  • Temperature range?  
  • Correlation with B (rotation) type?
  • Data logging?
  • Option items
     
    • Minimum volume for measurement?
    • Customized sensor design?
    • Battery model?
    • Unstable measurement
    • More about tighter measurement




    Q&A


    Why ultrasonic?

    Because our unique principle (velocity for concentration meter & torsion mode vibration for viscosity meter), our products do not require any periodical parts replacement.  Moreover, periodical calibration is not required.  This provides minimum running cost for the owner.

    How to purchase?

    Please refer to the following chart (concentration meter).
    First contact
    please inform us “the measurement subject name”, “the concentration range” & “the temperature range”.  In case of no measuring experience, we would like to practice sample testing for your subject (free service).

    Specification checkplease inform us the hardware specifications such as “sensor material”, “flange size”, “explosion proof model” & “cable length”.

    Consideration
    Business discussion/presentation by official agent & free demo trial is available.

    Order
    It takes approximately 6 to 8 weeks for delivery.  Please check the specific delivery date when order issue.

    Installation
    Supervising by our official agent & at-site technical support by our engineer (charged service) are available.

    More about sample testing

    In case of that your subject is now for us, we practice sample testing with free of charge.   Please ask our local agent for the detail.  In general, sample testing requires approximately 2 weeks.

    More about explosion proof model

    The explosion proof repeater is required for explosion proof model even though the distance is less than 10m.  Please review the specifications of each products for explosion proof specification.

    More about demo trial

    Free demo trial is available in order to confirm actual performance at site.  Please ask the local agent for the demo unit.  The trail period is 3 weeks in general.




    Concentration Meter (FUD-1 series)

    How to calibrate?

    Adjustment with OFFSET & GAIN functions is available in case of finding the gap between the concentration reading on display & absolute (analysis) value.

    How to check the condition

    Please prepare water in vessel, and measure velocity & temperature of water.  When the gap between the standard water velocity & the present velocity reading is less than 2m/sec, the product as hardware works normally.  Also, please check transducer condition such as scratch, dent & stain.
    Standard water velocity

    Any replacement item?

    This product does not feature any item which can ware out or age.  However, we suggest to replace a power supply unit for the product over 10 years.  Also, replacement of an explosion proof repeater over 15 years is suggested.

    Any caution at installation?

    The common trouble on our product is measurement error due to bubble affection.  Bubble interrupts or reduces the ultrasonic pulse travelling in the solution.  We strongly suggest to review bubble condition at the installation position in advance.
    Instruction guide for FUD-1

    Measurement errors

    ?Bubble affection
    Bubbles in solution can be critical for measurement.  Please avoid bubble affection absolutely.
    ?Stain / Sludge / accretion
    Stain, sludge & accretion should cause reduction of ultrasonic pulse.  Periodical cleaning is suggested under these conditions.
    ?Noise affection
    Noise typically causes spikes on reading.  Please give proper countermeasure such as “changing GND and/or power source route”, “Cable re-wiring” & “noise filter”.




    Viscosity Meter (FUV-1 series)


    How to calibrate?

    Adjustment with OFFSET & GAIN functions is available in case of finding the gap between the concentration reading on display & absolute value.

    How to check the condition?

    Please check the viscosity reading on display under condition of air (around 0 mPa.S), water (0 to 2 mPa.S) & JIS standard oil (minimum gap).


    Any replacement item?

    This product does not feature any item which can ware out or age.  However, we suggest to replace a power supply unit for the product over 10 years.  Also, replacement of an explosion proof repeater over 15 years is suggested.

    Measurement range?

    FUV-1 features 3 types based on viscosity range as low viscosity (0-100mPa.S), middle viscosity (0-1,000mPa.S) & high viscosity (0-10,000mPa.S).   Temperature limitation for low & middle is 0-100 degree C, and for high is 0-50 degree C.




    Desktop Viscosity Meter (FCV-100)


    Measurement range?

    FCV-100 features 3 types based on viscosity range as low viscosity (0-100mPa.S), middle viscosity (0-1,000mPa.S) & high viscosity (0-10,000mPa.S).

    Temperature range?

    10 to 40 degree C.


    Correlation with B (rotation) type?

    Certain correlation can be found with B-type for Newtonian fluid theoretically.  However, it is uncertain for Non-Newtonian fluid due to different shearing speed.  FCV-100 has viscosity conversion function (liner compensation with 10 points data input) for this requirement.

    Data logging?

    RS232C communication software (CD) is bundled.  It is possible to store the log data (as text file) in your laptop with this software.  Moreover, you can make a real time graph for viscosity & temperature.

    Option items

    Carrying case, cramp stand & RS232C cable are available as extra charged option.  Please ask the local agent in detail.

    Minimum volume for measurement?

    With the extra option of the small volume vessel, measurement is available for 30 ml.

    Customized sensor design?

    Not available.

    Battery model

    Not available.

    Unstable measurement

    The followings are typical conditions of unstable measurement.  Please avoid these conditions for proper measurement.
    1. Incomplete immersing
    2. Unstable temperature
    3. Touching with side wall of the vessel
    4. Stain / accretion on the transducer
    5. Ununiformed viscosity (incomplete mixture) of the sample

    More about tighter measurement

    • Stable temperature condition with temperature controller (water bath)
    • Warm up time (20 to 30 min.)
    • Position (measuring condition) maintaining with cramp holder

    FUJI ULTRASONIC ENGINEERING
    1068 Iida-cho, Chuo-ku, Hamamatsu, Shizuoka 435-0028
    TEL.+81-53-464-6460  FAX.+81-053-465-3815
    主站蜘蛛池模板: 毛片a级毛片免费播放100| 我和岳乱妇三级高清电影| 百合潮湿的欲望| 狠狠干2022| 精品人妻系列无码人妻漫画| 狠狠躁天天躁中文字幕无码| 欧美性猛交XXXX乱大交3| 日本成人福利视频| 小兔子好大从衣服里跳出来 | 最新国产精品视频| 成人麻豆日韩在无码视频| 夜夜橹橹网站夜夜橹橹| 国产无遮挡裸体免费视频| 嗯啊公交车上被群j| 亚洲欧美精品午睡沙发| 久久精品国产亚洲av麻豆 | 日本三浦理惠子中文字幕| 天天操天天操天天射| 国产欧美va欧美va香蕉在线观看 | 野花香高清在线观看视频播放免费| 精品一区二区三区无码视频| 绝世名器np嗯嗯哦哦粗| 欧美精品九九99久久在免费线| 波霸影院一区二区| 日韩精品久久无码中文字幕| 日本毛茸茸的丰满熟妇| 无码人妻久久一区二区三区不卡| 日韩在线视频二区| 无人区1080在线完整免费版| 国产精品香蕉在线一区| 国产a级黄色片| 亚洲国产一区二区a毛片| 一级毛片在线播放| 国产香蕉在线精彩视频| 特级aaaaaaaaa毛片免费视频| 日韩在线精品视频| 国内精品九九久久久精品| 国产99视频在线| 国产一级毛片网站| 亚洲日韩国产精品无码av| 中文字幕一区二区三区乱码|